3-22-2018
Abstract
Introduction: Computer-assisted navigation systems have been explored in total hip arthroplasty (THA) to improve component positioning. While these systems traditionally rely on anterior pelvic plane registration, variances in soft tissue thickness overlying anatomical landmarks can lead to registration error, and the supine coronal plane has instead been proposed. The purpose of this study was to evaluate the accuracy of a novel navigation tool, using registration of the anterior pelvic plane or supine coronal plane during simulated anterior THA.
Methods: Measurements regarding the acetabular component position, and changes in leg length and offset were recorded. Benchtop phantoms and target measurement values commonly seen in surgery were used for analysis. Measurements for anteversion and inclination, and changes in leg length and offset were recorded by the navigation tool and compared with the known target value of the simulation. Pearson's
Results: The device accurately measured cup position and leg length measurements to within 1° and 1 mm of the known target values, respectively. Across all simulations, there was a strong, positive relationship between values obtained by the device and the known target values (
Conclusion: The preliminary findings of this study suggest that the novel navigation tool tested is a potentially viable tool to improve the accuracy of component placement during THA using the anterior approach.
Recommended Citation
Parvizi, Javad MD; Benson, Jessica R.; and Muir, Jeffrey M., "A new mini-navigation tool allows accurate component placement during anterior total hip arthroplasty." (2018). Department of Orthopaedic Surgery Faculty Papers. Paper 114.
https://jdc.jefferson.edu/orthofp/114
Creative Commons License
This work is licensed under a Creative Commons Attribution-Noncommercial 3.0 License
PubMed ID
29606894
Language
English
Comments
This article has been peer reviewed. It is the author’s final published version in Medical Devices: Evidence and Research, Volume 11, March 2018, Pages 95-104.
The published version is available at https://doi.org/10.2147/MDER.S151835. Originally published in Dove Medical Press. Copyright © Parvizi et al.