Abstract
BACKGROUND: High-sensitivity cardiac troponin (hs-cTn) assays performed in the central laboratory have long turnaround times. This study evaluates the clinical performance of a novel bedside point-of-care (POC) hs-cTnI assay (i-STAT® hs-TnI) to aid in the diagnosis of myocardial infarction (MI).
METHODS: A prospective multisite cohort study involving emergency department (ED) patients with symptoms suggestive of acute coronary syndrome was conducted at 28 sites across the United States. Serial whole blood samples were collected at 0-1 h and >1-3 h from ED presentation. Primary outcome was MI or non-MI. Sensitivity, specificity, and negative predictive value (NPV) for MI were calculated with 97.5% lower confidence limits (LCL) using the female, male, and overall 99th percentile upper reference limit (URL, 13, 28, and 21 ng/L, respectively). Post hoc analysis was also conducted for initial measurements below the limit of quantification (LoQ, 2.9 ng/L) and < 5.0 ng/L.
RESULTS: The cohort included 3582 patients, median age 59 y, 64% female. MI incidence was 8.6%. Sensitivity and specificity at >1-3 h were 96.6% (LCL 91.7%) and 82.1% (80.2%) using the female URL; 90.7% (LCL 84.1%) and 83.9% (LCL 81.4%) using the male URL; and 92.4% (LCL 88.3%) and 85.6% (LCL 84.2%) using the overall URL. NPVs for MI of initial hs-cTnI measurements
CONCLUSIONS: The whole blood POC i-STAT hs-TnI assay demonstrated performance characteristics similar to other hs-cTn assays and achieved high sensitivity and NPV for initial measurements
CLINICALTRIALS.GOV REGISTRATION NUMBER: NCT05629572.
Recommended Citation
Wu, Alan H. B.; Mahler, Simon A.; House, Stacey L.; Jortani, Saeed A.; Creer, Michael H.; Rising, Kristin L.; Bischof, Jason J.; Crystal, Chad; Headden, Gary F.; Apple, Fred S.; Harrison, Nicholas E.; Meltzer, Andrew C.; Baca, Justin T.; Tiffany, Brian; Sochor, Mark; Pentecostes, Grace U.; Di Tullio, Katrina P.; Gupta, Manish S.; Singer, Adam J.; and Januzzi, James L., "Clinical Performance of a Novel Point-of-Care High-Sensitivity Cardiac Troponin I Assay" (2026). Department of Emergency Medicine Faculty Papers. Paper 283.
https://jdc.jefferson.edu/emfp/283
Creative Commons License

This work is licensed under a Creative Commons Attribution 4.0 License.
PubMed ID
41146629
Language
English
Included in
Emergency Medicine Commons, Health Services Administration Commons, Investigative Techniques Commons, Laboratory Medicine Commons


Comments
This article is the author’s final published version in The journal of applied laboratory medicine, Volume 11, Issue 1, 2026, Pages 98-112.
The published version is available at https://doi.org/10.1093/jalm/jfaf152. Copyright © 2024 Author(s) (or their employer(s)).